Štúdium možností prenosu a on-line chirálnej separácie vybraných pesticídov metódou HPLC s použitím techniky prepínania kolón

Autoři

  • M. Chalányová Katedra analytickej chémie, Prírodovedecká fakulta, Univerzita Komenského v Bratislave, Bratislava, Slovensko
  • I. Petránová Výskumný ústav vodného hospodárstva, Bratislava, Slovensko

Klíčová slova:

on-line prenos, chirálna HPLC separácia, pesticídy, prepínanie kolón

Abstrakt

The work deals with finding a suitable achiral stationary phase for preconcentration of pesticides epoxiconazole and permethrin. The break-through curves were measured for the studied analytes on several stationary phases. Sorption capacity of the sorbents showed that Silasorb Phenyl is suitable for the preconcentration of epoxiconazole and permethrin.

HPLC separation conditions in reverse-phase mode of analytes were optimized on the chiral stationary phase ChiraDex. For on-line transfer of epoxiconazol and permethrin on achiral column Silasorb Phenyl and following back-flush washout on chiral stationary phase ChiraDex column-switching technique has been used. The resolution of epoxiconazol enantiomers e1–e2 is 1.48; permethrin enantiomers p1–p2 is 1.2; and permethrin enantiomers p3–p4 is 2.2.

Stahování

Publikováno

15.03.2016

Jak citovat

Chalányová, M., & Petránová, I. (2016). Štúdium možností prenosu a on-line chirálnej separácie vybraných pesticídov metódou HPLC s použitím techniky prepínania kolón. Chemické Listy, 110(3), 185–189. Získáno z http://ww.chemicke-listy.cz/ojs3/index.php/chemicke-listy/article/view/214

Číslo

Sekce

Články

Nejaktuálnější články stejného autora (stejných autorů)